header background water

Equipment from 50 ml up to 500 ml

Description

The SFE is versatile "multi-solvent" extraction equipment with precise control of process parameters for analytical purposes :

  • Extraction with supercritical CO2 at pressures up to 1000 bar (15 000 PSI) and temperatures up to 150°C allowing very high pressure extraction over a very wide range of operating parameters and more opportunities to use supercritical CO2 as non-flammable, non-toxic and environment friendly solvent.
  • High pressure and subcritical water extraction up to 400 bar and 250°C.
laboratory equipment bench sfe

Include components :

Components Description
CO2 pump Variable flow rate from 0 to 100 g/min.
Variable pressure from 0 to 1000 bar.
Extration vessel 50 / 100 / 200 / 500 ml.
Oven Used to heat the extration vessel.
Separator For extract separation / collection of extracts and CO2 recycling.
Pressure Control Back pressure regulator to control the pressure.
Control board Display and record process value.

Option :

Components Description
Co-solvent pump Flow rate up to 24 ml/min and pressure at 400 bar.
Filter Absorbent bed cartridge in the recycling loop to trap lightest and volatile particules.
Stirrer Supply of a magnetic stirrer installed on the extractor at 350 or 700 bar.
Automation Data acquisition and recording process value on PC.
Additional control with automated valve, to be discussed.
Spare vessel Supply of additional spare vessel.
Sampling loop Pressurized sampling loop for collecting extract in a precise volume allowing to make on line sample analysis.
Column Liquid fractionation column module.

Image3D

laboratory equipment bench drawing 3d

SFE with Co-Solvent pump and stirrer

Other option :

Components Description
Sapphire Windows on the extraction cell for visualization (consult Extratex to define the conditions).
Spectroscopy On-line spectroscopy cell to be placed at the exit of the extractor.